Poojalabequips

Products

PLE 118 Vertical Lab Oven

Laboratory ovens provide uniform temperature and precise temperature control for heating, baking, evaporating, sterilizing and other industrial laboratory functions.

PLE 118 Vertical Lab Oven

Pooja Lab ovens are available in two versions i.e. non-GMP and GMP choose from four different sizes suitable for various applications in Clinical Laboratory, Pathological Laboratory, Hospitals, Pharmaceuticals, Industrial Chemicals, R & D Labs, and drying glassware and sterilizing.Our laboratory ovens are widely used for day to day heating and drying applications performed in various laboratories. 

Pooja Lab ovens are available in two versions i.e. non-GMP and GMP choose from four different sizes suitable for various applications in Clinical Laboratory, Pathological Laboratory, Hospitals, Pharmaceuticals, Industrial Chemicals, R & D Labs, and drying glassware and sterilizing.Our laboratory ovens are widely used for day to day heating and drying applications performed in various laboratories. We specialize in both standard and customized models, specifically designed to meet the challenging demands of various scientists for individual and specialized research applications. These ovens are energy efficient, long life and low maintenance. Temperature range of ovens are 50oC to 250o C ± 2o C.

Model No.

Capacity in liters /cu. ft.

No. Of trays

Interior dime. WxDxH (inches/mm

External dime. WxDxH (inches/mm)

OV-01

0600/21.08

2 Nos

44x18.0x46 (1118x0457x1168)

50x25.0x75 (1270x0635x1905)

OV-02

0817/28.65

2 Nos

44x22.5x50 (1118x0572x1270)

50x29.5x79 (1270x0749x2007)

OV-03

0947/33.23

2 Nos

44x22.5x58 (1118x0572x1473)

50x29.5x87 (1270x0749x2210)

OV-04

1335/46.82

2 Nos

62x22.5x58 (1575x0572x1473)

68x29.5x87 (1727x0749x2210)

PRE-INSTALLATION REQUIREMENT

Stabilised 230 V 50Hz AC single phase 20 amps power supply with MCB and proper earthing within the radius one meter from the installation. To be installed in an air-conditioned room or in a cross ventilated, dust free and clean environment with free space around (Room temperature should be below 30 deg. C.). Zero level flooring